Fast X-ray detector system with simultaneous measurement of timing and energy for a single photon
Fast X-ray detector system with simultaneous measurement of timing and energy for a single photon
We developed a fast X-ray detector system for nuclear resonant scattering (NRS) experiments. Our system employs silicon avalanche photo-diode (Si-APD) as a fast X-ray sensor. The system is able to acquire both timing and energy of a single X-ray photon simultaneously in a high rate condition, $10^6$ counts per second for one Si-APD. The performance of the system was investigated in SPring-8, a synchrotron radiation facility in Japan. Good time resolution of \SI{120}{ps} (FWHM) was achieved with a slight tail distribution in the time spectrum by a level of $10^{-9}$ at 1 ns apart from the peak. Using this system, we successfully observed the NRS from the 26.27-keV level of mercury-201, which has a half-life of $630 \pm 50$ ps. We also demonstrated the reduction of background events caused by radioactive decays in a radioactive sample by discriminating photon energy.
Y. Miyamoto、A. Yamaguchi、S. Kitao、T. Hiraki、K. Okai、M. Seto、H. Hara、T. Masuda、R. Ozaki、K. Yoshimura、S. Okubo、A. Yoshimi、S. Uetake、N. Sasao、Y. Yoda
粒子探测技术、辐射探测技术、核仪器仪表
Y. Miyamoto,A. Yamaguchi,S. Kitao,T. Hiraki,K. Okai,M. Seto,H. Hara,T. Masuda,R. Ozaki,K. Yoshimura,S. Okubo,A. Yoshimi,S. Uetake,N. Sasao,Y. Yoda.Fast X-ray detector system with simultaneous measurement of timing and energy for a single photon[EB/OL].(2017-02-15)[2025-07-20].https://arxiv.org/abs/1702.04651.点此复制
评论