|国家预印本平台
首页|Coulomb blockade thermometry based nanocalorimetry

Coulomb blockade thermometry based nanocalorimetry

Coulomb blockade thermometry based nanocalorimetry

来源:Arxiv_logoArxiv
英文摘要

Specific heat is a powerful probe offering insights into the entropy and excitation spectrum of the studied material. While it is well established, a key challenge remains the measurements of microcrystals or thin films especially in the sub-Kelvin, high magnetic field regime. Here we present a setup combining the high sensitivity of SiN$_x$ membrane based calorimetry with the absolute accuracy of Coulomb blockade thermometry to realise a nanocalorimeter for such tasks. The magnetic field independent technique of Coulomb blockade thermometry provides an on-platform thermometer combining a primary thermometry mode for in-situ calibration with a fast secondary mode suitable for specific heat measurements. The setup is validated using measurements of a 20 $μ$g sample of Sr$_3$Ru$_2$O$_7$ achieving a resolution on the order of 0.1 nJ/K at 500 mK and an absolute accuracy limited by the determination of the sample's mass. Measurements of CeRh$_2$As$_2$ further highlight the benefits of measuring microcrystals with such a device.

Mari C. Cole、Maximilian T. Pelly、Craig V. Topping、Thomas Reindl、Ulrike Waizmann、Jürgan Weis、Andreas W. Rost

10.1063/5.0257902

物理学材料科学

Mari C. Cole,Maximilian T. Pelly,Craig V. Topping,Thomas Reindl,Ulrike Waizmann,Jürgan Weis,Andreas W. Rost.Coulomb blockade thermometry based nanocalorimetry[EB/OL].(2025-07-15)[2025-07-25].https://arxiv.org/abs/2507.11327.点此复制

评论