首页|Neutron and photon detector responses outside the lateral shield of a 3 MeV ⁷Li(p,n) accelerator-based BNCT neutron source: a PHITS and Geant4 study

Neutron and photon detector responses outside the lateral shield of a 3 MeV ⁷Li(p,n) accelerator-based BNCT neutron source: a PHITS and Geant4 study

mukhtar, Mr. ameer Mukhtar, Mr. Ameer Li, Dr. Jie Li, Dr. Zhifeng Wang, Prof. Sheng

✕

Neutron and photon detector responses outside the lateral shield of a 3 MeV ⁷Li(p,n) accelerator-based BNCT neutron source: a PHITS and Geant4 study

Neutron and photon detector responses outside the lateral shield of a 3 MeV ?Li(p,n) accelerator-based BNCT neutron source: a PHITS and Geant4 study

mukhtar, Mr. ameer Mukhtar, Mr. Ameer 1Li, Dr. Jie 2Li, Dr. Zhifeng 1Wang, Prof. Sheng3

作者信息

  • 1. Xi'an Jiaotong University
  • 2. Xi'an Jiaotong University;Xi’an Jiaotong University
  • 3. RIKEN;Xi’an Jiao-tong University;XJTU-Huzhou Neutron Science Laboratory
  • 折叠

摘要

Monitors around an accelerator-based boron neutron capture therapy (AB-BNCT) source measure a mixed neutron-photon field that differs from their calibration fields. This work characterises the field outside the lateral shield of a 3.0 MeV, 10 mA ⁷Li(p,n) source and the detector readings it produces. An energy-angle correlated source was generated with PHITS 3.35 from 10⁸ protons in thick natural lithium with evaluated proton data. It gave 2.37 × 10⁻⁴ neutrons per proton (1.48 × 10¹³ n s⁻¹ at 10 mA), with a mean energy of 0.51 MeV and 40% of neutrons emitted between 60° and 120°. This source was transported through a 22 cm wall of borated polyethylene, lead and concrete. Ambient dose equivalent H*(10) was scored at three ring-shaped monitoring positions, and NaI(Tl) and moderated ³He-counter responses were computed with Geant4 and PHITS. The wall reduced the neutron H*(10) 672-fold and the photon H*(10) 490fold. Outside it, neutrons dominated the dose (5.0 times the photon dose), and streaming around the beam-port end raised the neutron dose 2.3-fold. A forward-only source underestimated the lateral neutron dose 1.9-2.3-fold. Analytical checks confirmed the simulations, and the two codes agreed within 0.6% on the NaI(Tl) response. Moderated ³He counters would record about 2 × 10⁵ counts s⁻¹. The calibration factor of small spheres changed up to 1.8-fold between the shielded and streaming-affected fields, whereas that of a 10-inch sphere was unchanged within its 7% uncertainty. Lateral-field assessments of ⁷Li(p,n) sources need a target-level source, explicit streaming treatment and large or field-calibrated monitors.

Abstract

Monitors around an accelerator-based boron neutron capture therapy (AB-BNCT) source measure a mixed neutron-photon field that differs from their calibration fields. This work characterises the field outside the lateral shield of a 3.0 MeV, 10 mA Li(p,n) source and the detector readings it produces. An energy-angle correlated source was generated with PHITS 3.35 from 10 protons in thick natural lithium with evaluated proton data. It gave 2.37 10 neutrons per proton (1.48 10 n s at 10 mA), with a mean energy of 0.51 MeV and 40% of neutrons emitted between 60 and 120. This source was transported through a 22 cm wall of borated polyethylene, lead and concrete. Ambient dose equivalent H*(10) was scored at three ring-shaped monitoring positions, and NaI(Tl) and moderated He-counter responses were computed with Geant4 and PHITS. The wall reduced the neutron H*(10) 672-fold and the photon H*(10) 490fold. Outside it, neutrons dominated the dose (5.0 times the photon dose), and streaming around the beam-port end raised the neutron dose 2.3-fold. A forward-only source underestimated the lateral neutron dose 1.9-2.3-fold. Analytical checks confirmed the simulations, and the two codes agreed within 0.6% on the NaI(Tl) response. Moderated He counters would record about 2 10 counts s. The calibration factor of small spheres changed up to 1.8-fold between the shielded and streaming-affected fields, whereas that of a 10-inch sphere was unchanged within its 7% uncertainty. Lateral-field assessments of Li(p,n) sources need a target-level source, explicit streaming treatment and large or field-calibrated monitors.

关键词

Accelerator-based BNCT/⁷Li(p/n) neutron source/PHITS/GEANT4/Ambient dose equivalent/NaI(Tl) spectrometer/Bonner sphere/Lateral shielding

引用本文复制引用

mukhtar, Mr. ameer,Mukhtar, Mr. Ameer,Li, Dr. Jie,Li, Dr. Zhifeng,Wang, Prof. Sheng.Neutron and photon detector responses outside the lateral shield of a 3 MeV ⁷Li(p,n) accelerator-based BNCT neutron source: a PHITS and Geant4 study[EB/OL].(2026-11-01)[2026-11-04].https://chinaxiv.org/abs/202611.00027.

学科分类

粒子探测技术、辐射探测技术、核仪器仪表
首发时间: 2026-11-01