|国家预印本平台
首页|上海光源BL02U2线站掠入射X射线衍射仪的研制

上海光源BL02U2线站掠入射X射线衍射仪的研制

evelopment of a grazing incidence x-ray diffractometer for BL02U2 at SSRF

中文摘要英文摘要

上海光源BL02U2线站已经安装了两台德国HUBER公司的衍射仪,为了进一步满足用户对掠入射衍射实验方法的需求,在两台衍射仪之间安装了一台新的掠入射衍射仪。本文介绍了掠入射衍射仪的整体方案,并针对高精度重载大行程位移台、探测器三维调节平台、样品台系统、控制系统等重要模块进行了详细的论述。实现了三台衍射仪的兼容运行,并以MoS2样品为例介绍了掠入射衍射仪的典型实验结果。目前该掠入射衍射仪已经稳定运行超过2年,完善了上海光源的掠入射衍射实验条件,为材料表面科学研究提供了更丰富的实验平台。

Background : Two x-ray diffractometers produced by HUBER company have been installed at BL02U2 of Shanghai Synchrotron Radiation Facility (SSRF). Purpose :?In order to further meet the needs of users for grazing incidence diffraction experimental methods, a new grazing incidence diffractometer was installed ?between?the?two HUBER diffractometers. Methods :?This paper mainly introduces the overall scheme of the grazing incidence diffractometer, and discusses in detail several important modules such as high-precision, high-load and large-stroke linear stage, three-dimensional adjustment platform for detector, sample stage system, and the control system. Results :?The compatible operation of three diffractometers was realized, and the typical experimental result of the grazing incidence diffractometer with MoS2?sample was introduced as an example. Conclusions :?The grazing incidence x-ray diffractometer developed has been operated more than two years, and expands the experimental platform for material surface science research at SSRF.

顾月良、李钊、阴广志、张兴民、朱佩平、廖可梁、薛志鹏、李晓龙、郑旭

10.12074/202309.00113V1

物理学晶体学材料科学

X射线衍射,同步辐射,精密仪器

X-ray diffraction Synchrotron radiation Precision instrument

顾月良,李钊,阴广志,张兴民,朱佩平,廖可梁,薛志鹏,李晓龙,郑旭.上海光源BL02U2线站掠入射X射线衍射仪的研制[EB/OL].(2023-09-06)[2025-08-02].https://chinaxiv.org/abs/202309.00113.点此复制